Show filters
3 Total Results
Displaying 1-3 of 3
Sort by:
Attacker Value
Unknown
CVE-2020-12036
Disclosure Date: June 29, 2020 (last updated February 21, 2025)
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
0
Attacker Value
Unknown
CVE-2020-12035
Disclosure Date: June 29, 2020 (last updated February 21, 2025)
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
0
Attacker Value
Unknown
CVE-2020-12037
Disclosure Date: June 29, 2020 (last updated February 21, 2025)
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
0