Unknown
CVE-2020-27264
CVE ID
AttackerKB requires a CVE ID in order to pull vulnerability data and references from the CVE list and the National Vulnerability Database. If available, please supply below:
Add References:
Unknown
(0 users assessed)Unknown
(0 users assessed)CVE-2020-27264
MITRE ATT&CK
Collection
Command and Control
Credential Access
Defense Evasion
Discovery
Execution
Exfiltration
Impact
Initial Access
Lateral Movement
Persistence
Privilege Escalation
Topic Tags
Description
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications use deterministic keys, which allows unauthenticated, physically proximate attackers to brute-force the keys via Bluetooth Low Energy.
Add Assessment
No one has assessed this topic. Be the first to add your voice to the community.
CVSS V3 Severity and Metrics
General Information
Vendors
- sooil
Products
- anydana-a firmware,
- anydana-i firmware,
- diabecare rs firmware
References
Miscellaneous
Additional Info
Technical Analysis
Report as Emergent Threat Response
Report as Exploited in the Wild
CVE ID