Attacker Value
Unknown
(0 users assessed)
Exploitability
Unknown
(0 users assessed)
User Interaction
None
Privileges Required
None
Attack Vector
Network
0

CVE-2021-37160

Disclosure Date: August 02, 2021
Add MITRE ATT&CK tactics and techniques that apply to this CVE.

Description

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

Add Assessment

No one has assessed this topic. Be the first to add your voice to the community.

CVSS V3 Severity and Metrics
Base Score:
9.8 Critical
Impact Score:
5.9
Exploitability Score:
3.9
Vector:
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector (AV):
Network
Attack Complexity (AC):
Low
Privileges Required (PR):
None
User Interaction (UI):
None
Scope (S):
Unchanged
Confidentiality (C):
High
Integrity (I):
High
Availability (A):
High

General Information