Attacker Value
Unknown
(0 users assessed)
Exploitability
Unknown
(0 users assessed)
User Interaction
Unknown
Privileges Required
Unknown
Attack Vector
Unknown
0

CVE-2014-5434

Disclosure Date: March 26, 2019
Add MITRE ATT&CK tactics and techniques that apply to this CVE.

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.