CVE-2019-13531

Disclosure Date: November 08, 2019
Add MITRE ATT&CK tactics and techniques that apply to this CVE.

Description

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.

Add Assessment

No one has assessed this topic. Be the first to add your voice to the community.

CVSS V3 Severity and Metrics
Base Score:
4.6 Medium
Impact Score:
3.6
Exploitability Score:
0.9
Vector:
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
Attack Vector (AV):
Physical
Attack Complexity (AC):
Low
Privileges Required (PR):
None
User Interaction (UI):
None
Scope (S):
Unchanged
Confidentiality (C):
None
Integrity (I):
High
Availability (A):
None

General Information

Additional Info

Technical Analysis