CVE-2015-3459

Disclosure Date: April 29, 2015
Add MITRE ATT&CK tactics and techniques that apply to this CVE.

Description

The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.