Show filters
7 Total Results
Displaying 1-7 of 7
Sort by:
Attacker Value
Unknown
CVE-2021-38398
Disclosure Date: September 30, 2021 (last updated November 28, 2024)
The affected device uses off-the-shelf software components that contain unpatched vulnerabilities. A malicious attacker with physical access to the affected device could exploit these vulnerabilities.
1
Attacker Value
Unknown
CVE-2021-38396
Disclosure Date: September 30, 2021 (last updated November 28, 2024)
The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.
0
Attacker Value
Unknown
CVE-2021-38392
Disclosure Date: September 30, 2021 (last updated November 28, 2024)
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
0
Attacker Value
Unknown
CVE-2021-38394
Disclosure Date: September 30, 2021 (last updated November 28, 2024)
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.
0
Attacker Value
Unknown
CVE-2021-38400
Disclosure Date: September 30, 2021 (last updated November 28, 2024)
An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password.
0
Attacker Value
Unknown
CVE-2017-14012
Disclosure Date: May 01, 2018 (last updated November 26, 2024)
Boston Scientific ZOOM LATITUDE PRM Model 3120 does not encrypt PHI at rest. CVSS v3 base score: 4.6; CVSS vector string: AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
0
Attacker Value
Unknown
CVE-2017-14014
Disclosure Date: May 01, 2018 (last updated November 26, 2024)
Boston Scientific ZOOM LATITUDE PRM Model 3120 uses a hard-coded cryptographic key to encrypt PHI prior to having it transferred to removable media. CVSS v3 base score: 4.6; CVSS vector string: AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
0