Disclosure Date: January 28, 2019 (last updated November 27, 2024)
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
0
Attacker Value
Unknown
CVE-2018-19014
Disclosure Date: January 28, 2019 (last updated November 27, 2024)
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.
0
Attacker Value
Unknown
CVE-2018-19012
Disclosure Date: January 28, 2019 (last updated November 27, 2024)
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system.