Show filters
2 Total Results
Displaying 1-2 of 2
Sort by:
Attacker Value
Unknown
CVE-2025-24836
Disclosure Date: February 13, 2025 (last updated February 14, 2025)
With a specially crafted Python script, an attacker could send
continuous startMeasurement commands over an unencrypted Bluetooth
connection to the affected device. This would prevent the device from
connecting to a clinician's app to take patient readings and ostensibly
flood it with requests, resulting in a denial-of-service condition.
0
Attacker Value
Unknown
CVE-2025-23421
Disclosure Date: February 13, 2025 (last updated February 14, 2025)
An attacker could obtain firmware files and reverse engineer their
intended use leading to loss of confidentiality and integrity of the
hardware devices enabled by the Qardio iOS and Android applications.
0