Show filters
2 Total Results
Displaying 1-2 of 2
Sort by:
Attacker Value
Unknown

CVE-2018-14781

Disclosure Date: August 13, 2018 (last updated November 27, 2024)
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
0
Attacker Value
Unknown

CVE-2018-10634

Disclosure Date: August 13, 2018 (last updated November 27, 2024)
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G communications between the pump and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.
0