Show filters
4 Total Results
Displaying 1-4 of 4
Sort by:
Attacker Value
Unknown

CVE-2018-18984

Disclosure Date: December 14, 2018 (last updated November 27, 2024)
Medtronic CareLink 2090 Programmer CareLink 9790 Programmer 29901 Encore Programmer, all versions, The affected products do not encrypt or do not sufficiently encrypt the following sensitive information while at rest PII and PHI.
Attacker Value
Unknown

CVE-2018-10596

Disclosure Date: July 03, 2018 (last updated November 26, 2024)
Medtronic 2090 CareLink Programmer all versions The affected product uses a virtual private network connection to securely download updates. The product does not verify it is still connected to this virtual private network before downloading updates. An attacker with local network access to the programmer could influence these communications.
0
Attacker Value
Unknown

CVE-2018-5448

Disclosure Date: May 04, 2018 (last updated November 26, 2024)
All versions of the Medtronic 2090 Carelink Programmer are affected by a directory traversal vulnerability where the product's software deployment network could allow an attacker to read files on the system.
0
Attacker Value
Unknown

CVE-2018-5446

Disclosure Date: May 04, 2018 (last updated November 26, 2024)
All versions of the Medtronic 2090 Carelink Programmer are affected by a per-product username and password that is stored in a recoverable format which could allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network.
0