Show filters
39 Total Results
Displaying 31-39 of 39
Sort by:
Attacker Value
Unknown

CVE-2020-12016

Disclosure Date: June 29, 2020 (last updated February 21, 2025)
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI.
Attacker Value
Unknown

CVE-2020-12048

Disclosure Date: June 29, 2020 (last updated February 21, 2025)
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Attacker Value
Unknown

CVE-2020-12041

Disclosure Date: June 29, 2020 (last updated February 21, 2025)
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
Attacker Value
Unknown

CVE-2020-12032

Disclosure Date: June 29, 2020 (last updated February 21, 2025)
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI.
Attacker Value
Unknown

CVE-2014-5431

Disclosure Date: March 26, 2019 (last updated November 27, 2024)
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
0
Attacker Value
Unknown

CVE-2014-5433

Disclosure Date: March 26, 2019 (last updated November 27, 2024)
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
0
Attacker Value
Unknown

CVE-2014-5432

Disclosure Date: March 26, 2019 (last updated November 27, 2024)
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
0
Attacker Value
Unknown

CVE-2014-5434

Disclosure Date: March 26, 2019 (last updated November 27, 2024)
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
0
Attacker Value
Unknown

CVE-2012-4493

Disclosure Date: November 02, 2012 (last updated October 05, 2023)
Cross-site scripting (XSS) vulnerability in the administrative interface in the Better Revisions module 7.x-1.x before 7.x-1.1 for Drupal allows remote authenticated users with the "administer better revisions" permission to inject arbitrary web script or HTML via unspecified vectors.
0