Show filters
11 Total Results
Displaying 1-10 of 11
Sort by:
Attacker Value
Unknown
CVE-2014-5401
Disclosure Date: March 26, 2019 (last updated November 27, 2024)
Hospira MedNet software version 5.8 and prior uses vulnerable versions of the JBoss Enterprise Application Platform software that may allow unauthenticated users to execute arbitrary code on the target system. Hospira has developed a new version of the MedNet software, MedNet 6.1. Existing versions of MedNet can be upgraded to MedNet 6.1.
0
Attacker Value
Unknown
CVE-2015-7909
Disclosure Date: January 22, 2016 (last updated November 25, 2024)
Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 Infusion System 13.40 allows remote attackers to cause a denial of service or possibly have unspecified other impact via traffic on TCP port 5000.
0
Attacker Value
Unknown
CVE-2015-1011
Disclosure Date: July 06, 2015 (last updated October 05, 2023)
Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors.
0
Attacker Value
Unknown
CVE-2015-3958
Disclosure Date: July 06, 2015 (last updated October 05, 2023)
Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets.
0
Attacker Value
Unknown
CVE-2014-5406
Disclosure Date: July 06, 2015 (last updated October 05, 2023)
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
0
Attacker Value
Unknown
CVE-2015-3957
Disclosure Date: July 06, 2015 (last updated October 05, 2023)
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.
0
Attacker Value
Unknown
CVE-2015-3955
Disclosure Date: July 06, 2015 (last updated October 05, 2023)
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.
0
Attacker Value
Unknown
CVE-2015-3459
Disclosure Date: April 29, 2015 (last updated October 05, 2023)
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
0
Attacker Value
Unknown
CVE-2014-5400
Disclosure Date: April 03, 2015 (last updated October 05, 2023)
The installation component in Hospira MedNet before 6.1 places cleartext credentials in configuration files, which allows local users to obtain sensitive information by reading a file.
0
Attacker Value
Unknown
CVE-2014-5403
Disclosure Date: April 03, 2015 (last updated October 05, 2023)
Hospira MedNet before 6.1 uses hardcoded cryptographic keys for protection of data transmission from infusion pumps, which allows remote attackers to obtain sensitive information by sniffing the network.
0